• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: POSEY PRODUCTS POSEY BED 8070; PATIENT BED WITH CANOPY/RESTRAINTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

POSEY PRODUCTS POSEY BED 8070; PATIENT BED WITH CANOPY/RESTRAINTS Back to Search Results
Model Number 8070
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: evaluation of the returned product found one element damaged on the molding zipper (a patient access area) which could allow the unintentional release of a patient.Applying pressure directly to the unsecured area will reveal the breach, which is why the user manual advises caregivers to apply direct pressure along the entire length of the zipper.Posey beds are multi-use, serviceable items.As such, it is anticipated that the units may occasionally require repair, and as part of standard care the beds should be inspected prior to use.If damage is noted during these routine bed inspections, the unit should not be put into use with a patient and should be returned to posey for servicing.Instructions for use (ifu) were reviewed and found to provide adequate instructions and warnings for safe and effective use of the device.Service issues are trended and reviewed by management on a monthly basis.As part of this monthly review, trends and excursions above control limits will be assessed, documented and acted upon as warranted.No corrective or preventative actions are necessary at this time.(b)(4).
 
Event Description
Customer reported via email there is a broken zipper.Limited information provided.The date the issue was discovered is unknown and no patient incident or injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POSEY BED 8070
Type of Device
PATIENT BED WITH CANOPY/RESTRAINTS
Manufacturer (Section D)
POSEY PRODUCTS
2530 lindsay privado drive
unit a
ontario CA 91761
MDR Report Key12528140
MDR Text Key280537006
Report Number2020362-2021-00060
Device Sequence Number1
Product Code OYS
Combination Product (y/n)N
PMA/PMN Number
K113357
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial
Report Date 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number8070
Device Catalogue Number8070
Device Lot Number2020012826
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/29/2021
Initial Date FDA Received09/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-