Device Problem Computer Software Problem (1112)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/27/2021
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.(b)(4).The distributor performed a checkout of the equipment and confirmed the reported complaint.The carrier com express board was replaced to resolve the issue.Legal manufacturer: (b)(4).
 
Event Description
The hospital reported a software watchdog failure preventing mechanical ventilation.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 620
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
MDR Report Key12529698
MDR Text Key281608052
Report Number9710602-2021-00297
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial
Report Date 09/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/27/2021
Initial Date FDA Received09/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-