Catalog Number 425-00 |
Device Problem
Complete Loss of Power (4015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
The complaint is reported as: the unit "shuts off for no reason".The issue was detected prior to use on a patient.No patient involvement reported.
|
|
Event Description
|
The complaint is reported as: the unit "shuts off for no reason".The issue was detected prior to use on a patient.No patient involvement reported.
|
|
Manufacturer Narrative
|
Qn# (b)(4).The sample was returned for evaluation.A visual exam was performed and there were no major defects observed.Functional testing was performed and the unit was connected to 120vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was able to stay powered on and power off with no issues.A device history record review was performed and no relevant findings were identified.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
|
|