Catalog Number 81201 |
Device Problems
Complete Blockage (1094); Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.Hp had a kinked water hose causing no water flow to the insert.
|
|
Event Description
|
Based on the repair evaluation of the g130 scaler, the handpiece had a kinked water hose causing no water flow to the insert, no injury resulted.
|
|