Device evaluated by mfr: the device was returned to for analysis.The product return contained an angiojet solent dista.The assembly, supply line, shaft, tip, and spike line were all visually examined for potential damage.The device showed a hypotube fracture 25.5 cm from the tip.A functional test was performed.The pump was placed into the ultra drive console and the device displayed an under pressure error message.The device was unable to run, due to the pressure being below specifications.No further damage or discrepancies were noted on the rest of the catheter upon visual inspection.
|
Reportable based on the device analysis completed on 02-sep-2021.It was reported that error message occurred.An angiojet solent dista was used for thrombectomy.The catheter failed to prime throughout the procedure, and the device requested that the angiojet system be replaced with another catheter in order to continue working normally.The procedure was completed with another of the same device.There were no patient complications reported.However, returned device analysis revealed hypotube fracture.
|