MAUDE Adverse Event Report: BONALIVE BIOMATERIALS LTD BONALIVE GRANULES; BONE MATRIX IMPLANT, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BONALIVE BIOMATERIALS LTD BONALIVE GRANULES; BONE MATRIX IMPLANT, SYNTHETIC Back to Search Results
Catalog Number 13XXX
Device Problems Migration or Expulsion of Device (1395); Osseointegration Problem (3003); Insufficient Information (3190)
Patient Problems Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 03/09/2021
Event Type  Injury  
Manufacturer Narrative
The reason for granules moving / migrating is not known, but one possible option is challenges related to the operation.Ifu states in surgical procedure notes: "prevention of movement and migration of bonalive granules is essential for proper bone formation." and "it is important to have sufficient pre-existing bone wall in the defect and an adequate closure of the graft site (e.G.With cortical bone window, periosteum, collagen membrane, fascia or muscle flap)." and "the above surgical procedure notes are necessary for achieving good application performance and accurate obliteration, and to initiate the bioactive reaction." movement and migration, in turn, are the most probable reasons for inadequate bone formation.Two different products with different granules sizes has been used (2019 and 2021).The causal relationship between the device and the incident is not likely, but it is possible.Note: the 510k is for to catalog number 14xxx.The incident however, concerns catalog number 13xxx, which is not available in us and thus does not have a 510k.However, 13xxx and 14xxx are the same bioactive glass.
 
Event Description
There is only limited information available.Patient operated twice with bonalive granules (2019 and 2021).Maxillary cyst was removed and filled with bonalive granules.Almost a year after the first operation a granule came out from the upper jaw (complaint reported to manufacturer 22.10.2020).At that time further information from the patient and surgeon was asked, but not received.Incident report was not compiled.New information was received on 9.9.2021 from patient's family member: he reported that a reoperation was made where smaller bonalive granules were implanted.Now 6 months later granule begin to come out again and maxilla has not ossified.Additional information has been asked.
 
Manufacturer Narrative
Several contact attempts were made towards the complainant and the surgeon, but without success.Additional information was not received.The case remains unresolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONALIVE GRANULES
Type of Device
BONE MATRIX IMPLANT, SYNTHETIC
Manufacturer (Section D)
BONALIVE BIOMATERIALS LTD
biolinja 12
turku, 20750
FI  20750
Manufacturer (Section G)
BONALIVE BIOMATERIALS LTS
biolinja 12
turku, 20750
FI   20750
Manufacturer Contact
hanna ryosa
biolinja 12
turku, 20750
FI   20750
MDR Report Key12544562
MDR Text Key273739593
Report Number3008328709-2021-00002
Device Sequence Number1
Product Code MQV
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K191274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number13XXX
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/28/2021
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/17/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient Age61 YR
Patient SexFemale
-
-