The device was returned to philips.Visual inspection of the cable found that there was damage to the external jacket of the cable.This is a mechanical defect from misuse at customer site.It is believed that the damage is the result of the sensor cable being pinched locally (just the insulation, not the full cable).Since the opening is quite small, it is difficult to detect when checking the cable before use.At the same time, since the opening is quite small, there should not be a real problem with that cable because it does not allow direct contact between internal shield and patient skin." the functional testing of this cable was performed using the philips intellivue mx550 patient monitor along with the philips nmt patient module.The cable was connected however the inop message "nmt leads off" was produced.There was no observed heating along the patient cable.The cable was scrapped in the lab and no further action will be taken with this sample.The customer reported the patient burn was treated with dressing and neosporin.No further actions were taken and none are warranted.
|