According to the complaint the customer complains about deviation on the parameter ctbil during a patient measurement on a abl800 flex analyzer.The patient measurement sample #(b)(6) (17.09.2021; 10:10 p.M.) measured the value ctbil=10,8 mg/dl.A comparison measurement carried out at the customer's lab on a roche c702 cobas 8000 (16.09.2021; 22:20 p.M.) measured the value ctbil=21,58 mg/dl.Based on the abl800 flex analyzer measurement and the comparison measurement, the customer considers the 10,8 mg/dl ctbil measurement as false low.No reports of death or serious injury.
|