The user facility reported that they observed a small chemical burn on the patient's gastrointestinal tract during the procedure which resulted in a procedure delay.No treatment was administered.The procedure was completed successfully.Following notification of the reported event, a steris account manager spoke with user facility personnel and learned that the scope used during the procedure had been manually cleaned with revital-ox resert high level disinfectant and dried with non-steris alcohol.Based on the description of the event and the discussion with user facility personnel, it is likely that they did not properly rinse and dry the scope prior to use.The revital-ox resert high level disinfectant instructions for use include the following directions, "following removal from revital-ox resert high level disinfectant solution, thoroughly rinse the medical device by immersing it completely in water.Keep the instrument or medical device immersed for a minimum of 1 minute in duration, unless longer is specified by the instrument manufacturer.Manually flush all lumens with large volumes of water (at minimum 100 ml) unless otherwise noted by the instrument manufacturer.Remove device and discard rinse water.Do not reuse the water for rinsing or any other purpose.A final rinse using a 70% isopropyl alcohol solution (followed by medical air purge if within lumens) may be used to speed the drying process.Follow facility policies.Consult the instrument manufacturer's instructions for recommended rinsing and drying procedure." steris offered an in-service training on the proper use of revital-ox resert high level disinfectant, specifically, properly rinsing and drying of medical devices; however, the user facility declined.A 3-year complaint review indicates this to be an isolated event.No additional issues have been reported.
|