Catalog Number 4628 |
Device Problem
Device Damaged Prior to Use (2284)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference (b)(4).
|
|
Event Description
|
It was reported that, during wound treatment, it was impossible to take the dressing out from the pouch of a opsite flexigrid 6x7cm ctn 100 because the film adhered to the inside the pouch.It was found in 2 dressings in the carton.Treatment was resumed, without any delay, with a back-up device.Patient was not injured as consequence of this problem.
|
|
Manufacturer Narrative
|
Section: h3, h6: the batch record was reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of the finished product specification.There was also no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.A complaint history review found a small number of similar instances of the reported event.There is nothing to indicate that this is outside of acceptable rates of occurrence.The device was used for treatment and was not returned for analysis.We have not been able to confirm a relationship between the event and the device or identify a definitive root cause.It was reported that the dressing could not be used as it had stuck to the pouch.Probable root cause is an environmental issue such as temperature.Heat can alter the adhesive nature of the dressing, which may cause the adhesive to leak and stick to the pouch.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
|
|
Manufacturer Narrative
|
It was reported that the dressing could not be used as it had stuck to the pouch.The device was used for treatment and was returned for analysis.This was found to be as described in the complaint.This and comprehensive photographs provided by the customer were able to confirm a relationship between the event and the device the batch record was reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of the finished product specification.There was also no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.A complaint history review found no similar instances of the reported event in the last 3 years.The risk files, mitigate the reported issues with no updates required.Probable root cause is a missed quality check.The dressing is not aligned with the backing paper which has caused it to stick to the pouch.This faulty dressing should have been rejected and replaced with a new dressing.Strict checks are in place to make sure that faulty dressings are removed and do not leave manufacture.The investigation is now complete with no corrective actions required.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
|
|
Search Alerts/Recalls
|