• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. OXF ANAT BRG RT MD SIZE 4 PMA; OXFORD HXLPE BEARINGS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF ANAT BRG RT MD SIZE 4 PMA; OXFORD HXLPE BEARINGS Back to Search Results
Model Number N/A
Device Problems Break (1069); Fracture (1260); Material Integrity Problem (2978)
Patient Problem Insufficient Information (4580)
Event Date 08/01/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).Initial report.Report source, foreign - event occurred in (b)(6).Concomitant medical products: customer has indicated that the product will not be returned to zimmer biomet for investigation, as its location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the oxford bearing fractured.No further information in available at this time.
 
Event Description
It was reported that the oxford bearing fractured.No further information in available at this time.
 
Manufacturer Narrative
(b)(4).This final report is being submitted to relay additional information.Complaint summary: adequate photographs have not been provided and product has not been returned for evaluation.Therefore, the investigation has been limited to the information provided, a review of the device history records, and a complaint history search.Review of the device history records identified no deviations or anomalies during manufacturing that could be related to the reported event.Review of complaint history identified no additional similar complaints for the reported item and no additional complaints for the reported item 650-1055 and lot 3062797 combination.These devices are used for treatment.The reported event is not related to a combination of products; therefore, a compatibility review is not applicable.These part and lot combinations are not associated with any recalls at the time the search was conducted.The likely condition of the devices when it left zimmer biomet is conforming to specification.The reported event has not been confirmed as relevant photographs have not been provided and packaging has not been returned for evaluation and the dhr review did not identify any issues.The root cause of the reported event cannot be determined with the information provided.Corrective or preventative action not required.The root cause of the reported event cannot be determined with the information provided.The investigation is completed based on current available information.If any additional information becomes available, then the complaint will be reopened and further investigated.H3 other text : device location unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF ANAT BRG RT MD SIZE 4 PMA
Type of Device
OXFORD HXLPE BEARINGS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12553274
MDR Text Key274001649
Report Number3002806535-2021-00424
Device Sequence Number1
Product Code NRA
UDI-Device Identifier05019279786220
UDI-Public05019279786220
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/11/2012
Device Model NumberN/A
Device Catalogue Number159576
Device Lot Number1291789
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/21/2021
Initial Date FDA Received09/30/2021
Supplement Dates Manufacturer Received01/11/2022
Supplement Dates FDA Received02/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/13/2007
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age72 YR
Patient SexMale
-
-