Model Number G31525
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2021
Event Type  malfunction  
Event Description
Cook fusion cytology brush was used to brush biliary duct to check for abnormal cells.The flexible tip of the brush broke off in the patients biliary duct.Brush was pulled out of scope and sent to cytology.Retrieval balloon was then used to go inside biliary duct and sweep out tip of brush.Forceps were used to grab the tip and pull it out of patient.Tip of brush was then packaged in its original package and sent to device manufacturer to be examined.Physician then did another sweep of the duct and did a cholangiogram to ensure there was no perforation of the duct.Manufacturer response for fusion cytology brush, (brand not provided) (per site reporter): unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION
Type of Device
ENDOSCOPIC CYTOLOGY BRUSH
Manufacturer (Section D)
COOK INCORPORATED
750 daniels way
bloomington IN 47404
MDR Report Key12553823
MDR Text Key274011089
Report Number12553823
Device Sequence Number1
Product Code FDX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG31525
Device Catalogue NumberFS-CB-1.5-S
Device Lot NumberW4476479
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA09/10/2021
Event Location Hospital
Date Report to Manufacturer09/30/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age26280 DA
Patient Weight66
-
-