EDWARDS LIFESCIENCES DR DISPOSABLE PRESSURE TRANSDUCER KIT; TRANSDUCER, PRESSURE, CATHETER TIP
|
Back to Search Results |
|
Model Number PXVFL21049 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not returned for evaluation, since it was discarded at the facility.Without return of the unit it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.It is not known if some procedural factors may have contributed to the event.No corrective actions will be taken at this time.The lot number was not provided thus a device history record was not reviewed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
|
Event Description
|
It was reported that inaccurate arterial blood pressure value was observed on the first day of use.The patient was not treated based on the incorrect value.Dpt zeroed before use.The data log was not provided.Dpt zeroing was performed again, but the problem was not solved, so the device was exchanged and the problem was solved.A philips patient monitor was used.No error message was displayed on the monitor.Patient demographic information requested but unavailable.The device was discarded by the hospital due to infection.There were no patient complications reported.
|
|