Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at olympus service operation repair center (sorc) on september 7, 2021, it was found that the angle was locked due to wear of the angle wire.There was no report of patient injury associated with the event.
|
The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.The manufacturing record was reviewed and found no irregularities.Omsc presumed that the angle wire wore out because a load was applied during angulation, and the wire was caught in the cable support and the operation deteriorated.It was also presumed that since the instrument channel had pinhole according to the evaluation information from sorc, the operation deteriorated due to rusting of the wire and curved tube caused by water immersion.
|