Catalog Number 00625006535 |
Device Problem
Fracture (1260)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 09/09/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation as the product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during a hip procedure the surgeon screwed the trilogy self-tapping screw in to the acetabulum and part of the screw head fractured and remains in patient.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Surgical note provided state that small crack was seen during the time of implantation, however a small corner of the screw was taken out and the screw was placed in the poly.X-ray images were not sent to mmi for the fractured screw left in and is not visible because it's obscured by the cup and would not enhance the investigation.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|