MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II XENON LIGHT SOURCE Back to Search Results
Model Number CLV-180
Device Problems Defective Component (2292); Installation-Related Problem (2965)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/10/2021
Event Type  malfunction  
Event Description
A user facility reported to olympus that the evis exera ii xenon light source display blurry image.The event occurred during procedure.A similar device was used to complete the procedure.There was no patient harm or user injury reported due to the event.During a standard service inspection of the customer returned device, improper lamp installation was noted.This report is to capture the reportable malfunction found at estimation.There was no patient injury or harm reported.
 
Manufacturer Narrative
The device was returned for investigation.Upon evaluation of the device, the lamp heat-sink, lamp shaft, sink spring, and front pole were burned.In addition, a non-olympus lamp life meter reading over 500+hours, light output measured out of range, and worn out socket slider switch caused intermittent use of high intensity mode were note.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the event of a blurry image and the lamp heat sink, lamp shaft, sink spring, front pole being burned could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12564480
MDR Text Key280933506
Report Number8010047-2021-12585
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170215544
UDI-Public04953170215544
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 11/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/27/2021
Initial Date FDA Received10/01/2021
Supplement Dates Manufacturer Received11/05/2021
Supplement Dates FDA Received11/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/26/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-