Model Number CLV-S190 |
Device Problem
Energy Output Problem (1431)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device was returned to local service department of olympus.Local service department checked the subject device and found that the reported phenomenon could be duplicated due to power supply unit failure.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that during the preparation for use, the lamp didn't work.The facility finished the intended case with another similar device.There was no report of patient injury associated with the event.The subject device was used in combination with cv-190.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The case is not due to design.It has been over 15 years since the subject device was manufactured.Based on the results of the investigation, the cause of the faulty switch regulator could not be identified.The faulty switch regulator was replaced; as a result, the device was inspected and passed all olympus's functional tests.Olympus will continue to monitor.
|
|