Event Description
During a cystoscopy and tur for bladder.The ge uroview x-ray malfunctioned.The foot pedal stuck and was beeping.It was frozen.The engineer checked the system, rebooted, relay replaced, and indicator light on wall in operating room not functioning, replacement light bulb ordered.Fda safety report id # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO TABLE WITH FLUORO
Type of Device
SYSTEM, X-RAY, FLUOROSCOPIC, IMAGE-INTENSIFIED
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS, INC.
MDR Report Key12568834
MDR Text Key274769142
Report NumberMW5104339
Device Sequence Number1
Product Code JAA
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCI-K5-160-TS
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/01/2021
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age54 YR
Patient Weight95
-
-