HILL-ROM BATESVILLE VERSACARE FRAME; BED, FLOTATION THERAPY, POWERED Back to Search Results
Model Number P3201
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2019
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the brake alarm was not sounding when the brake was released.Per the hillrom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Make sure the brake is not set, and then plug the bed into an applicable power source.A steady alarm comes on.Set the brake.The alarm stops.Repair or replace as necessary.As no serial number was provided for this bed, a search of the hillrom maintenance records for any hillrom performed preventative maintenance on this bed was unable to be completed.It is unknown if the facility performs preventative maintenance on their beds.Hillrom has contacted the account requesting repair of the bed with this alleged issue.The account is scrapping the bed due to age and refused for a repair.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the bed had no audible brake not set alarm.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE FRAME
Type of Device
BED, FLOTATION THERAPY, POWERED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
harish vishwanathan
1069 state route 46 east
batesville, IN 47006
3127289851
MDR Report Key12573048
MDR Text Key274795715
Report Number1824206-2021-00492
Device Sequence Number1
Product Code IOQ
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP3201
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/01/2019
Initial Date FDA Received10/04/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-