Model Number 25-40-45 |
Device Problem
Gas Output Problem (1266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and the error did not occur at the device startup.The device could not be tested on site due to gas supplies at the hospital being in use.The unit will be returned at the manufacturer site for further analysis.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 gas blender system gave an error code associated to a discrepancy between the set and the actual gas flow values during service.There was no report of patient injury.
|
|
Manufacturer Narrative
|
H.10: the affected device was returned to the manufacturer site for repair.The reported failure was traced back to a defective mass flow controller for o2.The component will be replaced to solve the reported issue.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: the mass flow controller for o2 was replaced and the issue could be solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
|
|
Event Description
|
See initial report.
|
|