LIVANOVA DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-40-45
Device Problem Gas Output Problem (1266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2021
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and the error did not occur at the device startup.The device could not be tested on site due to gas supplies at the hospital being in use.The unit will be returned at the manufacturer site for further analysis.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 gas blender system gave an error code associated to a discrepancy between the set and the actual gas flow values during service.There was no report of patient injury.
 
Manufacturer Narrative
H.10: the affected device was returned to the manufacturer site for repair.The reported failure was traced back to a defective mass flow controller for o2.The component will be replaced to solve the reported issue.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: the mass flow controller for o2 was replaced and the issue could be solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key12582689
MDR Text Key274954392
Report Number9611109-2021-00559
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-40-45
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/06/2021
Initial Date FDA Received10/06/2021
Supplement Dates Manufacturer Received11/22/2021
03/30/2022
Supplement Dates FDA Received12/22/2021
04/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-