Catalog Number 2C4711K
Device Problem Backflow (1064)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/09/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a backflow leak was observed from the fill port of a small volume folfusor.The backflow out of the fill port was observed while filling the device with medication prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the actual device was received for evaluation.Visual inspection on the returned device via the naked eye noted evidence of leak at the fillport when the fillport cap was removed.Further inspection on the device revealed the cause of backflow was due to a particle measured to be 2.37 mm in length, stuck under the device checkband.The particle was identified to be acrylic material via fourier transform infrared spectroscopy (ftir) test.Acrylic is the material of the device stressmember.The reported condition was verified.The cause of the material under the checkband could not be determined; however, the most probable cause is related to small shaving getting stuck under the checkband during manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key12601543
MDR Text Key276118499
Report Number1416980-2021-06163
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412082172
UDI-Public(01)00085412082172
Combination Product (y/n)Y
Reporter Country CodeFR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4711K
Device Lot Number21C027
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/14/2021
Initial Date FDA Received10/08/2021
Supplement Dates Manufacturer Received11/01/2021
Supplement Dates FDA Received11/10/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-