Model Number CLV-190
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2021
Event Type  malfunction  
Manufacturer Narrative
The suspect device was returned to olympus and evaluated.The reported problem could not be duplicated or confirmed.No sign of burning was noted and the fan was verified to function as expected.No problems with power were observed.The investigation is ongoing and the root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
A user facility reported to olympus that a bright light was observed then smoke.The problem, as reported to olympus, was first identified during preparation for use.The intended procedure was a colonoscopy and was completed after a delay of approximately 30 minutes using different equipment.There was no patient injury and no injury to the end users, associated with the problem, reported to olympus.This is report #2 of 2 due to multiple devices involved.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.Based on the results of the investigation, the specific root cause could not be determined at this time as the olympus service center was unable to reproduce the bright light and smoke from the device.User was using a non-olympus lamp which life meter reading was out of range at 469 hours and 50 minutes.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12602084
MDR Text Key275837541
Report Number8010047-2021-12942
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 11/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/29/2021
Initial Date FDA Received10/09/2021
Supplement Dates Manufacturer Received11/05/2021
Supplement Dates FDA Received11/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/10/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OLYMPUS, MODEL CV-190.
-
-