• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TNDM BP SHL/XLPE LNR 51OD 26ID; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TNDM BP SHL/XLPE LNR 51OD 26ID; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Catalog Number 71325051
Device Problem Patient Device Interaction Problem (4001)
Patient Problems Pain (1994); Joint Dislocation (2374)
Event Date 09/28/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that, after a partial hip replacement surgery was performed on (b)(6) 2021, the patient had to undergo a revision surgery on (b)(6) 2021 due to pain, loosening, and dislocation.Patient outcome remains unknown.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, this case reports a revision due to pain, loosening, and dislocation within 2 weeks of implantation of a partial hip replacement.Patient current health status is unknown.Requested information beyond the revision surgery was not provided.Without clinically relevant information the root cause of the reported revision could not be concluded.The patient impact was pain, loosening and a revision surgery.Further patient impact could not be assessed.No further clinical/medical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, joint tightness, material in use, patient reaction, patient anatomy, abnormal loading of limb, abnormal motion over time, bone degeneration, fit/sizing issue or loss of sterility.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TNDM BP SHL/XLPE LNR 51OD 26ID
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12603043
MDR Text Key276584099
Report Number1020279-2021-07497
Device Sequence Number1
Product Code JDG
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71325051
Device Lot Number19DM25210
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/28/2021
Initial Date FDA Received10/10/2021
Supplement Dates Manufacturer Received11/16/2021
Supplement Dates FDA Received11/19/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/25/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
71305414-SYN SEL II POR SO 10/12 SZ 14
Patient Outcome(s) Other;
-
-