• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN 6 DEG VALG COL NTRAL S3-8; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN 6 DEG VALG COL NTRAL S3-8; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71434404
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that, during a tka surgery, a legion 6 deg valgus collet neutral s3-8 no longer attached to necessary instruments.Surgery was resumed, without any delay, with a back-up device.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN 6 DEG VALG COL NTRAL S3-8
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12606102
MDR Text Key276160268
Report Number1020279-2021-07506
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010550170
UDI-Public03596010550170
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71434404
Device Catalogue Number71434404
Device Lot Number06BM05264
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/28/2021
Initial Date FDA Received10/11/2021
Supplement Dates Manufacturer Received09/28/2021
Supplement Dates FDA Received11/12/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-