• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST PST; PROTHROMBIN TIME TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST PST; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 07762798003
Device Problem Low Readings (2460)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/08/2021
Event Type  malfunction  
Manufacturer Narrative
The returned test strips were measured on reference meters with a high-level control sample: all inr values were within the specified target ranges, confirming the functionality of the complained coaguchek test strips.Test 1: 2.9 inr, test 2: 2.8 inr, test 3: 2.9 inr.No error messages occurred.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.The returned customer material and retention material complied with the specification.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods." iniital reporter occupation was lay user/patient.
 
Event Description
There was an allegation of a questionable result from a coaguchek inrange meter.The serial number was requested but was not provided.In the morning at an unknown time, the result from a laboratory using an unknown reagent was 2.7 inr.At 7:15 am, the result from the meter was 1.6 inr.The therapeutic range was 2.0-2.5 inr.The testing frequency was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST PST
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key12617600
MDR Text Key276147785
Report Number1823260-2021-02962
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K170960
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number07762798003
Device Lot Number46588512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2021
Initial Date Manufacturer Received 09/20/2021
Initial Date FDA Received10/12/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age60 YR
Patient Weight96
-
-