During an atrial fibrillation procedure, validation issues occurred due to an issue with the surface link causing a delay.When the patch was applied and validated, it succeeded, but an alert was issued requesting revalidation.Upon confirmation, the body surface electrocardiogram signal was not output to ensitex.Reconnecting the electrocardiogram cable and system patch, power cycling, and re-pasting the system reference were performed with no resolution.The study was completed using the ensite precision in the hospital as an alternative device.The procedure was delayed by more than 50 minutes.In the verification after the procedure, it was confirmed that the pin of the system reference patch insertion port was scraped.The patch and electrocardiograms were connected to this surface link, but the body surface electrocardiogram was not displayed.When connected to a new alternative surface link, a body surface signal was displayed and was therefore determined that the surface link did not function as expected.
|
Additional information: g3, h2, h3
the results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received and images provided the reported event can be confirmed.
|