ENCORE MEDICAL L.P DISCOVERY ELBOW; DISC ULNA BEARING KIT Back to Search Results
Catalog Number 114800
Device Problems Break (1069); Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 09/17/2021
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event Description
Revision surgery - bearing screw was broken and replaced poly and bearing kit.
 
Manufacturer Narrative
The reason for this revision surgery was reported as broken bearing screw.The actual length of in-vivo for the items listed is unknown as the original surgery date was not provided or could be established.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.This investigation is limited in scope as only partial information was provided to djo surgical for review.The revised items were not returned for examination and the lot numbers were not provided.Given the limited information, a search of djo and available zimmer biomet records for the previous surgery produced no results, therefore; the items could not be identified.To adequately investigate this event, the lot numbers are necessary.If this information is submitted at a future date, this investigation will be re-evaluated.Customer complaint history of the reported items showed no present trends or on-going issues that are needing a review.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The surgeon performed this revision to remedy the patient's condition.This complaint will be closed pending receipt of additional information.No further action is deemed necessary at this time.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOVERY ELBOW
Type of Device
DISC ULNA BEARING KIT
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
kiersten soderman
9800 metric blvd
austin, TX 78758-5445
MDR Report Key12619014
MDR Text Key276094156
Report Number1644408-2021-01082
Device Sequence Number1
Product Code JDC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K013042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number114800
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/17/2021
Initial Date FDA Received10/12/2021
Supplement Dates Manufacturer Received11/18/2021
Supplement Dates FDA Received12/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
540-00-000 LOT UNKNOWN
Patient Outcome(s) Required Intervention;
Patient Age64 YR
Patient SexMale
-
-