(b)(4).Method: the mr290v vented autofeed humidification chamber was returned to fisher & paykel healthcare (f&p) in new zealand where it was visually inspected and analysed.Results: visual inspection of the returned mr290v chamber revealed a horizontal crack line on the chamber dome below the maximum water line.It was also observed that there were smudge marks on the chamber dome near the location of the chamber crack indicating that the chamber was in contact with a solution containing alcohol.Further thermoanalytical testing also revealed that the chamber dome showed extensive gouging/scuffing near the fracture point of the chamber crack which is consistent with externally applied forces causing the stress damage.Conclusion: we are unable to determine the root cause of the chamber crack, however our investigation indicates that the crack is most likely due to mechanical stress.The stress source was unable to be identified.It should be noted that based on previous investigations of the mr290v chamber, exposure to alcohol based chemicals may affect the integrity of the chamber dome.This can potentially lead to stress cracking of the chamber dome in combination with externally applied forces.The reported patient respiratory distress occurred while the patient was provided manual bagging with a non-f&p device.The patient further desaturated to 60%-70% spo2 when they were placed on the new set up.The patient was provided "manual breaths" via the ventilator and recovered to 94%-97% spo2.Every mr290 chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber dome due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290v chamber would have met the required specification at the time of production.Our user instructions that accompany the mr290v vented autofeed humidification chamber state the following: - "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions, could lead to a loss of ventilation pressure" - "do not use the chamber if the seals are not intact when received, or if it has been dropped." -"the use of breathing circuits, chambers, accessories, or combinations which are not approved by fisher & paykel healthcare may result in poor humidification system performance, ventilator malfunction, and harm to the patient/ user".
|