Model Number IPN913776 |
Device Problem
Complete Loss of Power (4015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
|
Event Description
|
It was reported "broken unit.The heater turns on but doesn't stay on".No injury or harm reported.Patient condition reported as "fine".
|
|
Event Description
|
It was reported "broken unit.The heater turns on but doesn't stay on".No injury or harm reported.Patient condition reported as "fine".
|
|
Manufacturer Narrative
|
(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no major defects observed.Functional testing was performed and the unit was connected to 120vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was able to stay powered on for approximately 30 minutes with no issues.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
|
|
Search Alerts/Recalls
|