MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DXTEND GLENOSPHERE STD D42MM; DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DXTEND GLENOSPHERE STD D42MM; DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE Back to Search Results
Model Number 130760142
Device Problems Difficult to Insert (1316); Device Damaged Prior to Use (2284)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2021
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that this was an unknown procedure performed on (b)(6) 2021.The glenosphere was not able to be tightened to the metaglene (unk).It was found that the screw component on the glenosphere had been damaged during its insertion.There was no surgical delay and no harm to the patient.The devices were brand new and the first use when the issue occurred.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND GLENOSPHERE STD D42MM
Type of Device
DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12640057
MDR Text Key276685635
Report Number1818910-2021-22807
Device Sequence Number1
Product Code PHX
UDI-Device Identifier10603295027775
UDI-Public10603295027775
Combination Product (y/n)N
PMA/PMN Number
K120174
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number130760142
Device Catalogue Number130760142
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/04/2021
Initial Date FDA Received10/15/2021
Supplement Dates Manufacturer Received12/23/2021
Supplement Dates FDA Received12/29/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN SHOULDER METAGLENE
-
-