Model Number ADDR01 |
Device Problem
Failure to Interrogate (1332)
|
Patient Problem
Burn(s) (1757)
|
Event Date 09/29/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that, during the implantable pulse generator (ipg) changeout procedure, the patient experienced inadvertent atrial cauterization.It was also reported the implantable pulse generator (ipg) exhibited a telemetry problem, when tested intra-operatively, post use of cautery.The ipg was explanted.It was also reported that the right atrial (ra) lead exhibited high thresholds and loss of capture, when tested intra-operatively, post use of cautery.The ra lead was capped and replaced.On an attempt to replace the ra lead, the sheath could not be advanced past a superior vena cava (svc) narrowing, bending the existing ra lead and rv lead, but subsequently passed using a dilator.The ultimate replacement ra lead was capped for potential future use.The rv lead was capped for potential future use.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.The battery indicator signifying that it is time for device replacement occurred after explant.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that no malfunction was noted with the introducer sheath.The ipg was replaced.
|
|
Search Alerts/Recalls
|
|
|