Model Number 10-88-60
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/17/2021
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 mast roller pump gave a motor control failure error messages during priming.There was no report of patient injury.
 
Manufacturer Narrative
The customer originally provided the wrong serial number of the affected pump.The correct serial number has been added in the dedicated d.4 section and h.4 has been accordingly updated with the manufacturing date.H.10: a livanova field service technician was dispatched to the customer facility to investigate the device.The serial read-out (real time device parameters and setting recording file) was gathered and analyzed: the reported issue was confirmed.An error code associated to motor control failure was stored in the pump micro-controller.During the service activity the processor board of the control panel and internal cables were replaced but the error persisted.Thus, the motor control board and the motor power amplifier board were replaced and the problem could be solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Based on findings, the cause of the reported error message could be traced back to a defective motor control/motor power amplifier boards.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 MAST ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key12646071
MDR Text Key276862278
Report Number9611109-2021-00585
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K062396
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-88-60
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/20/2021
Initial Date FDA Received10/18/2021
Supplement Dates Manufacturer Received11/22/2021
Supplement Dates FDA Received12/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-