BD MEDICAL (BD WEST) MEDICAL SURGICAL 10 ML BD POSIFLUSH NORMAL SALINE SYRINGE; SALINE, VASCULAR ACCESS FLUSH Back to Search Results
Model Number 306546
Device Problem Complete Blockage (1094)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/20/2021
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Initial reporter facility name: fms himg dialysis center jv, fresenius medical care investigation summary: it was reported that a posiflush syringe will not push the solution out.To aid in the investigation, two samples with no packaging blisters were received for evaluation by our quality team.A visual inspection was performed and no defects or imperfections were observed.Each sample was then tested for sustaining force, which is the force applied to the plunger rod while moving downwards when expelling the saline solution, and all results were within specification.It could be possible that syringes that are towards the high specification limit are related to the symptom reported by the customer since they require extra force than normal to expel the solution.A device history record review was completed for provided material number 306546, lot number 1145120.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.Previous investigations have revealed that improper silicone application within the syringe barrel can create plunger resistance.Several quality initiatives have been implemented on our manufacturing line to ensure that the silicone application is properly applied during the manufacturing process.Additionally, a monitoring program is also in place to verify the silicone is uniformly applied to the syringe barrel.Based on the investigation and with the returned sample analysis the symptom reported by the customer could not be confirmed.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that the 10 ml bd posiflush¿ normal saline syringe experienced difficult plunger movement.The following information was provided by the initial reporter: a cm reported a posiflush pre-filled syringe will not push the solution out.It starts and then locks up about half way and will not go any farther, they have the syringe with nothing left in it.There was no patient harm or adverse event reported at intake.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10 ML BD POSIFLUSH NORMAL SALINE SYRINGE
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12648354
MDR Text Key276906529
Report Number1911916-2021-01081
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public30382903065463
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2024
Device Model Number306546
Device Catalogue Number306546
Device Lot Number1145120
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/20/2021
Initial Date FDA Received10/18/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-