Catalog Number 8065741115 |
Device Problems
Loose or Intermittent Connection (1371); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A physician reported the infusion tubings fell off easily when being connected to the phacoemulsification hand piece during a cataract procedure because of a relatively smaller white connector.The tubing was replaced with an alternate one and the procedure was completed.There was no harm to the patient.This is the first of three reports from this facility.
|
|
Manufacturer Narrative
|
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.No sample has been returned for evaluation; therefore, the condition of the product could not be verified.Visual inspection or functional testing could not be conducted in order to ascertain the failure mode for the consumable device.The root cause of the customer's complaint could not be established as a sample has not been received and the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|
Manufacturer Narrative
|
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.Based on our current tracking, there are no adverse trends for this reported complaint.The infusion tube was reported to be loose.Two multipak manifolds were visually inspected.A lab stock fluid management system (fms) was used to perform functional testing.No defects were found when connecting the connectors to cassette housing.The samples were tested using an infiniti console.The samples primed and tuned with the ultrasonic handpiece successfully.No message code, no fluid or air leaks, and no cracks on the luers were found.No occlusion was found in all manifolds.The root cause of the customer's complaint could not be established; the returned sample functioned per specification.This complaint has been reviewed and it is determined that no further actions will be pursed at this time as the sample met specifications.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|