OLYMPUS MEDICAL SYSTEMS CORP. FORCEPS/IRRIGATION PLUG (ISOLATED TYPE); FORCEPS/ IRRIGATION PLUG Back to Search Results
Model Number MAJ-891
Device Problem Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2021
Event Type  malfunction  
Manufacturer Narrative
In speaking with olympus technical support via the phone, the olympus sales representative was advised the customer would need to order a replacement.It was reported the subject device would not be returned for evaluation, therefore the device evaluation could not be completed at this time.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The olympus representative reported on behalf of the customer, during preparation for use for a diagnostic cystoscopy procedure, the forceps/irrigation plug (isolated type) was unable to be reassembled.The threaded side would not go together.The intended procedure was completed with another forceps/irrigation plug with no surgical delay.No patient harm reported.The customer had used the item for several months.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The dhr was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The case is not due to design.Based on the results of the investigation, it is likely that one of tightening rings, biopsy ports, or tightening ring attachment part of housing was damaged or foreign material adhered to any of the above contact surfaces during assembly.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORCEPS/IRRIGATION PLUG (ISOLATED TYPE)
Type of Device
FORCEPS/ IRRIGATION PLUG
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12655976
MDR Text Key281271397
Report Number8010047-2021-13328
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170063114
UDI-Public04953170063114
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K912120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAJ-891
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/06/2021
Initial Date FDA Received10/19/2021
Supplement Dates Manufacturer Received01/13/2022
Supplement Dates FDA Received01/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-