Catalog Number 388.720
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/20/2021
Event Type  malfunction  
Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from (b)(6) reports an event as follows: it was reported that during a procedure on (b)(6) 2021, the blade of the rod cutter broke.This occurred while cutting a cocr rod.No fragments came in contact with the patient.Surgery was completed with no delay.Patient outcome was good and as intended.This report is for a rod cutter.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: part number: 388.720, lot number: t197214, manufacturing site: tuttlingen, release to warehouse date: may 19, 2020, june 5, 2020.A review of the device history records was performed for the finished device lot number and a nonconformance (nc) was started because small pores were visible in the weld seam on a few devices and were reworked.The necessary actions to ensure the final product quality have been taken and documented in the appropriate quality system.The nc have no impact to the complaint condition.The final quality release criteria were met before this batch was released for distribution.The raw material certificate was reviewed and the used material was according to the specification of the device.Visual inspection: the bolt cutter (part# 388.720, lot# t197214) was returned and received at us customer quality (cq).Upon visual inspection, the cutter blade was observed to be broken.No other issues were identified with the returned device.Device failure/defect identified? yes.Dimensional inspection: dimensional inspection of the relevant feature of the received device was not performed at cq due to it being post manufacturing damage.Document/specification review: the following drawing(s) was reviewed: -rod cutter for rods up to 6mm diameter: current and manufactured no design issues or discrepancies were found during this investigation.Complaint confirmed? yes.Conclusion: the complaint condition was confirmed for the bolt cutter (part# 388.720, lot# t197214).No definitive root-cause can be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROD CUTTER
Type of Device
CUTTER, WIRE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen 78532
GM   78532
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12658877
MDR Text Key277445360
Report Number8030965-2021-08791
Device Sequence Number1
Product Code HXZ
UDI-Device Identifier07611819710504
UDI-Public(01)07611819710504
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number388.720
Device Lot NumberT197214
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/22/2021
Initial Date FDA Received10/19/2021
Supplement Dates Manufacturer Received11/19/2021
Supplement Dates FDA Received12/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK - RODS
-
-