MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION LITHOVUE; URETEROSCOPE AND ACCESSORIES, FLEXIBLE/RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION LITHOVUE; URETEROSCOPE AND ACCESSORIES, FLEXIBLE/RIGID Back to Search Results
Model Number 791-100
Device Problems No Display/Image (1183); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/24/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported to boston scientific corporation that a lithovue workstation was opened and used on (b)(6) 2021 in a ureteroscopy procedure.During the procedure, it was noted the plug receptacle on the monitor was loose which caused the connection to be lost.There was no other scope available at that time.The procedure was cancelled due to this event.There were no patient complications as a result of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LITHOVUE
Type of Device
URETEROSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue
attn: return products
st. paul MN 55112
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key12664147
MDR Text Key277380609
Report Number3005099803-2021-05359
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K153049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number791-100
Device Catalogue Number791-100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/27/2021
Initial Date FDA Received10/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-