The subject device has not been returned to omsc but was returned to olympus (b)(4) for evaluation.Olympus (b)(4) checked the subject device.It was found some black particles inside the disinfectant tank and disinfectant tank hose 1, of the subject device.Its tank hose must have been replaced with new one as repair.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.Based on the report of olympus china and it report said there was a need to replacement of tank hose 1 and the former similar case, omsc presumed there was the possibility this phenomenon was attributed to the following; the inner surface of the hose might have peeled off due to deterioration.When the disinfectant solution was pumped up, it might have appeared as the foreign substances and was reprocessed with those foreign substances.Therefore, it might have been mixed with the disinfectant solution.If additional information becomes available, this report will be supplemented.
|