Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at olympus service operation repair center (sorc) on september 29, 2021, it was found that the angulation of the subject device was locked due to corrosion of the angle mechanism of the subject device.Other detailed information was not provided.There was no report of patient injury associated with the event.This device is an olympus asset.
|
Olympus medical systems corp.(omsc) could not investigate the subject device, because the subject device was not returned to omsc.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.Since the subject device was not returned to omsc, the exact cause was unknown.Olympus service operation repair center (sorc) found that there was the air leak of the subject device.Omsc surmised that the reported phenomenon occurred since the angle mechanism of the subject device was corroded due to the air leak of the subject device.
|