U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC XR FP TYPE1 PPS 7X134MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC XR FP TYPE1 PPS 7X134MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported during incoming inspection at one of the warehouses, debris was found in the sterile packaging.No patient involvement.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No additional information on the reported event.
 
Manufacturer Narrative
The reported event is confirmed.Visual evaluation of the returned product found foreign debris is present inside the sterile packaging, and the sterile packaging remains sealed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR FP TYPE1 PPS 7X134MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12671562
MDR Text Key277627870
Report Number0001825034-2021-02900
Device Sequence Number1
Product Code LPH
UDI-Device Identifier00880304516908
UDI-Public(01)00880304516908(17)310516(10)7024014
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-102070
Device Lot Number7024014
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/28/2021
Initial Date FDA Received10/21/2021
Supplement Dates Manufacturer Received01/04/2022
Supplement Dates FDA Received01/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberNI
Patient Sequence Number1
-
-