Model Number N/A |
Device Problems
Tear, Rip or Hole in Device Packaging (2385); Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the implant was improperly packaged and the sterile package was damaged causing a 30 min surgical delay.No adverse events have been reported as a result of the malfunction.Additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual review of the device packaging shows the stem had punctured through both the inner cavity plastic, inner foam, outer carton and clear shrink film.The sterility has been breached.Dhr was reviewed and no discrepancies were found.The condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event can be attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|