The electrode belt sn (b)(4) was returned to the distributor and found to be fully functional.There is no indication of a device malfunction.Monitor sn (b)(4) was returned and evaluated at the distributor in accordance with procedures recommended by zoll manufacturing corporation.The reported problem (code 104) was confirmed.The cause for the failure was an ejected sd card.Upon investigation the monitor ecg acquisition and pulse delivery circuitry were tested and found to be fully functional.The root cause of the ejected sd card could not be positively identified.An ejected sd card does not affect the monitor's ability to detect and treat an arrhythmia.Unavailability of retrospective ecg recordings did not cause or contribute to the adverse event.The sd card is a data logger that only stores the patient's retrospective continuous ecg recordings.There is no real and active patient monitoring associated with the stored retrospective ecg recordings.
|