Model Number UNKNOWN
Device Problem Decrease in Suction (1146)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/24/2021
Event Type  malfunction  
Manufacturer Narrative
Expiration date: unknown as product lot number was not provided.Udi #: a complete udi # is unknown as product lot number was not provided.A partial number has been provided device manufacture date: unknown, as the lot number of the device was not provided reporter's email address: unknown/not provided.At the time of the investigation, product was not available for evaluation as the surgical procedure was completed with the same product.Therefore, no testing could be performed.The reported event cannot be confirmed.For this reported event, lot# was not reported and cannot be obtained.Therefore, manufacturing record review cannot be performed.Based on the information obtained, product malfunction and product deficiency cannot be confirmed.No further investigation is required.The patient interface (pi) suction ring may lose suction during a procedure.Label copy states corneal fixation vacuum loss can occur.There are several factors that may contribute to suction issues such as doctor¿s technique in applying the suction ring to the cornea, doctor¿s technique in squeezing the pi clip to secure the suction ring to the pi cone and patient anatomy affecting the interface between the patient¿s cornea and the suction ring.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported by the customer that they experienced suction loss with the patient interface(pi) during the laser firing.Procedure was completed successfully with the same pi.There was no patient injury or surgical intervention needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FS DISPOSABLE INTERFACE
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
MDR Report Key12677699
MDR Text Key278152240
Report Number2020664-2021-07814
Device Sequence Number1
Product Code HNO
UDI-Device Identifier15050474534688
UDI-Public(01)15050474534688
Combination Product (y/n)N
PMA/PMN Number
K060372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial
Report Date 10/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNKNOWN
Device Catalogue Number590106AN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/28/2021
Initial Date FDA Received10/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-