Model Number CLV-S45 |
Device Problem
Energy Output Problem (1431)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/29/2021 |
Event Type
malfunction
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that during the laryngoscopy, the ignitor was malfunctioning and the lamp didn't light up.The device was switched on and off repeatedly to solve the fault.The laryngoscopy was completed.There was no report of patient injury associated with the event.The subject device was used in combination with otv-s7.
|
|
Manufacturer Narrative
|
The subject device was returned to local service department of olympus.Local service department checked the subject device and found that the reported phenomenon could be duplicated due to power supply unit failure.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
|
|
Manufacturer Narrative
|
This is a supplemental report to provide additional information.The manufacturing record was reviewed and found no irregularities.Based on the inspection result by local service department, it was presumed that the reported phenomenon occurred due to converter malfunction.The exact cause of the reported event could not be conclusively determined.
|
|