• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VISERA XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VISERA XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S45
Device Problem Energy Output Problem (1431)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2021
Event Type  malfunction  
Event Description
Olympus medical systems corp.(omsc) was informed that during the laryngoscopy, the ignitor was malfunctioning and the lamp didn't light up.The device was switched on and off repeatedly to solve the fault.The laryngoscopy was completed.There was no report of patient injury associated with the event.The subject device was used in combination with otv-s7.
 
Manufacturer Narrative
The subject device was returned to local service department of olympus.Local service department checked the subject device and found that the reported phenomenon could be duplicated due to power supply unit failure.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Manufacturer Narrative
This is a supplemental report to provide additional information.The manufacturing record was reviewed and found no irregularities.Based on the inspection result by local service department, it was presumed that the reported phenomenon occurred due to converter malfunction.The exact cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12677954
MDR Text Key281852293
Report Number8010047-2021-13530
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170275777
UDI-Public04953170275777
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/29/2021
Initial Date FDA Received10/21/2021
Supplement Dates Manufacturer Received11/11/2021
Supplement Dates FDA Received11/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-