• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VISERA 4K UHD XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VISERA 4K UHD XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S400
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to olympus for evaluation.Olympus medical systems corp.(omsc) could not investigate the subject device, because the subject device was not returned to omsc.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.Since the subject device was not returned to omsc, the exact cause was unknown.Omsc surmised that the reported phenomenon occurred since the lamp has deteriorated and burned out due to repeated use for a long period of time.
 
Event Description
Olympus medical systems corp.(omsc) was informed that the local field service engineer checked the subject device at the facility and found that the examination lamp of the subject device was exhausted.The local field service engineer replaced the examination lamp with a new one.The event date was unknown.Other detailed information was not provided.There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA 4K UHD XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12678274
MDR Text Key280941322
Report Number8010047-2021-13538
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
PMA/PMN Number
K151011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 10/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCLV-S400
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/17/2021
Initial Date FDA Received10/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-