• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CLEARSTREAM TECHNOLOGIES LTD. BANTAM OTW PTA CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARSTREAM TECHNOLOGIES LTD. BANTAM OTW PTA CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number 48515030
Device Problems Difficult to Flush (1251); Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2021
Event Type  malfunction  
Manufacturer Narrative
Manufacturing review: a complaint history review was performed.This is the first complaint reported for this lot number.A device history record review was performed and the lot met all release criteria.Investigation summary: the device was returned for evaluation.The result of the investigation is confirmed for the reported flushing issue.The inner lumen of the device was attempted to be flushed however this was unsuccessful.The definitive root cause for the reported difficult to flush issue could not be determined based upon the available information.Labeling review: the instructions for use for the bantam pta balloon catheter was reviewed and contains the following information relevant to the reported event: directions for use: inspection and preparation remove the protective sheath by first withdrawing the stylet and then slowly removing the sheath while holding the catheter as close to the balloon as possible.If any resistance is felt, or if any stretching of the catheter is observed while removing the protective sheath, the product should not be used.The catheter should then be inspected for bends, kinks or stretched portions.Do not use if product damage is evident.Prepare a mixture of contrast medium and normal saline as per normal procedure.(recommended 25% /75%) attach a stopcock and a 20 ml syringe half filled with the contrast solution to the balloon port.Point the syringe nozzle downward and aspirate until all air is removed from the balloon.Turn the stopcock off and maintain the vacuum in the balloon.Purge the catheter guidewire lumen thoroughly.Reinserting the balloon into the protective sheath may damage the balloon or catheter.(expiry date: 04/2024).
 
Event Description
It was reported that during preparation for an angioplasty procedure, the pta catheter allegedly had flushing problem.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BANTAM OTW PTA CATHETER
Type of Device
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI  N A
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12679209
MDR Text Key277919216
Report Number9616666-2021-00172
Device Sequence Number1
Product Code LIT
UDI-Device Identifier05391515540507
UDI-Public(01)05391515540507
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K072947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number48515030
Device Lot NumberCMFS0083
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/14/2021
Initial Date FDA Received10/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
Patient Weight60
-
-