• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT 5F VASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT 5F VASCULAR STENT Back to Search Results
Catalog Number 5F061501C
Device Problems Break (1069); Entrapment of Device (1212); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2021
Event Type  malfunction  
Event Description
It was reported that during a stent placement procedure to treat the lesion in right superficial femoral artery using the ipsilateral approach, the delivery system allegedly got stuck near the middle of the stent.It was further reported that tip was allegedly detached and remained in the vessel and was subsequently retrieved using snare.There was no reported patient injury.
 
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the lifestent 5f vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent 5f vascular stent are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 12/2023).
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent 5f vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent 5f vascular stent are identified in d2 and g4.H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned for evaluation.Images/dicoms demonstrating the entrapment were not provided so that the alleged entrapment cannot be confirmed.The returned sample was found with activated system and without stent since the stent had been successfully deployed.The sample confirmed that the inner catheter cardan tube was broken towards the distal end which leads to confirmed evaluation result for catheter break.The stent was released with a 0.014" guidewire without problem, the lesion was not pre dilated and the stent was not properly aligned to the vessel wall directly after deployment.The definitive root cause could not be determined based upon available information.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instructions for use state: 'do not constrict the delivery system during stent deployment.If excessive force is felt during stent deployment, do not force the stent system.Remove the stent system and replace with a new unit.' in regards to pre dilation the instructions for use state: 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' under materials required the instructions for use state: '5f (1.67 mm) or larger introducer sheath (¿) 0.014 inch (0.36 mm) - 0.035 inch (0.89 mm) diameter guidewire'.In regards to removal difficulty the instructions for use state: 'if resistance is met while retracting the delivery system over a guidewire, remove the delivery system and guidewire together.' holding and handling of the system throughout deployment was found sufficiently described.H10: (expiry date: 12/2023).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a stent placement procedure in the right superficial femoral artery using the ipsilateral approach, after successful stent deployment, the delivery system allegedly got struck near the middle of the stent upon removal.It was further reported that tip was allegedly detached.Reportedly, the tip was retrieved using snare.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT 5F VASCULAR STENT
Type of Device
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12680969
MDR Text Key281221465
Report Number9681442-2021-00565
Device Sequence Number1
Product Code NIP
UDI-Device Identifier00801741120022
UDI-Public(01)00801741120022
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P070014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5F061501C
Device Lot NumberANEZ0430
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/24/2021
Initial Date FDA Received10/22/2021
Supplement Dates Manufacturer Received02/09/2022
Supplement Dates FDA Received02/17/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-