• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. UNKNOWN ONQ PUMP WITH CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. UNKNOWN ONQ PUMP WITH CATHETERS Back to Search Results
Model Number UNKNOWN
Device Problem Material Fragmentation (1261)
Patient Problems Pulmonary Embolism (1498); Foreign Body In Patient (2687); Thrombosis/Thrombus (4440); Peripheral Edema (4578)
Event Date 09/22/2021
Event Type  Injury  
Event Description
It was reported by the patient's daughter that following her mother's hysterotomy, she received a pain pump with two catheters placed at the surgical site.Upon removal several days later, one of the catheters broke, leaving a piece behind inside the patient.Based on comparison with the other catheter, she believes at least 5 inches remain in the patient's abdomen.A ct with contrast was performed, however, the fragment could not be visualized.Per additional information received 08 oct 2021, the patient's daughter reports that her mother received a total abdominal hysterectomy on (b)(6) 2021 at houston methodist sugar land hospital and was discharged with an onq pump on (b)(6) 2021.They were told that the catheters could be removed on (b)(6) 2021, however they were provided no instructions on how to do so by the hospital.Upon removal, one broke inside the patient, the other was removed without incident.The patient returned to the hospital on (b)(6) 2021 for "two pulmonary embolisms and severe swelling in arm, and blood clots," however she does not know if those incidents were related to the catheter break.The patient was treated with blood thinners and "other meds" and released home.At the time of this call, the patient was recovering at home.
 
Manufacturer Narrative
The sample is reported to be available, but has not yet been received by the manufacturer.Photos of the device were provided by the patient's mother.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 21 oct 2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).
 
Manufacturer Narrative
A catheter with missing black tip was returned for evaluation.Visual examination revealed that the catheter was stretched in appearance, starting at 48cms.The complaint is confirmed as reported, however, no root cause could be determined.All information reasonably known as of 16 nov 2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN ONQ PUMP WITH CATHETERS
Type of Device
CATHETERS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S. DE R.L. DE C.V.
ave noruega edificio d-1b
fraccionamiento rubio
tijuana b.c.
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key12682358
MDR Text Key281281448
Report Number2026095-2021-00100
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberUNKNOWN
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/06/2021
Initial Date FDA Received10/22/2021
Supplement Dates Manufacturer Received10/25/2021
Supplement Dates FDA Received11/17/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age73 YR
Patient SexFemale
-
-