Model Number XOM UNK MICDEBBUR |
Device Problems
Break (1069); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/30/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The product analysis result confirmed that the broken bur was attached and could not be removed.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported by health care professional that the bur could not be removed from the handpiece during the procedure.There was no patient or staff impact.S<(>&<)>r found that the bur was broken.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
On follow up it was reported that there was no fragments came off.The issue was discovered when the surgeon attempted to remove the bur.A back up device was used to complete the procedure.
|
|
Manufacturer Narrative
|
H3 the product analysis result indicates there was residue consistent with biological contaminants on the device.: visually, the inner shaft was bent until broken approximately 0.71¿ from the distal end of the inner hub.There were tool marks on the inner hub.There was deformation in the distal outside diameter of the inner hub.The outside diameter of the inner hub shall be 0.330¿ +/-0.002 and measured 0.328¿ in the undamaged area and 0.339¿ in the deformed area, in which the latter was out of specification.The inner hub bushing was rough and worn.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|