Model Number CDS0701-XTW |
Device Problem
Difficult to Flush (1251)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is filed to report difficult to flush.It was reported that during preparation of the clip delivery system (cds), there was no saline flushing through the sleeve.The stopcocks and lines were changed out, but there was no flow of the saline throughout the cds.Therefore, the cds was not used in the patient and the procedure was successfully completed with a new cds.There were no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from the lot.Based on available information, a cause for the reported difficulty flushing the clip delivery system could not be determined.There is no indication of product issue with respect to manufacture, design or labeling.
|
|